SIL Home
SIL On-line Manual
SIL Personnel
Stanford Stable Isotope Lab
SIL Analyses
Funding Support
Stanford University Home Page
GES Home Page
Other Links
Stanford University
Geological and Environmental Sciences, Bldg. 320
Stanford, CA 94305-2115
Tel: 650-723-0817, Fax:650-725-0979
Finnigan MAT Kiel III carbonate device interfaced with a MAT 252 IRMS:
Description: Individual acid bath carbonate system capable of analyzing 46 samples in an automated run.
Power requirements = 230V (-10%, + 6%), 16A three phase, 50/60Hz. Max permissible voltage between ground and neutral lines < 400mV.
Liquid nitrogen = 0.5 l/sample.
Phosphoric acid = 500ml reservoir.
Compressed air supply = 50 to 75psig. (350 to 500Kpa).
Compressed helium = 15ml/sample, industrial grade He.
Ambient temperature range = 64 to 82°F (18-28°C).
Relative humidity range = 20 to 70%.
Dimensions = 6.25ft H x 2.95ft W x 2.95ft D (190cm H x 90cm W x 90cm D).
Weight = 220lbs (100kg).
Kiel III connected to 252 IRMS specs:
External precision 1s @ 20µg = 0.03ppt 13C, 0.06ppt 18O.
Kiel III specs :
Sample amount = 20 - 100µg
Acid used = orthoH3PO4 @ 70°C, , 2 to 3 drops/sample
Sample capacity = 46 sample continuous automated run
Analysis time: d13C and d 18O approximately 30 minutes/sample