Finnigan MAT 252 Isotope Ratio Mass Spectrometer (IRMS):

 Description: The Finnigan MAT 252 IRMS has a dual micro-volume inlet system, electromagnet, 24X multiport manifold interfaced with a Pentium PC using ISODATTM software.

 Power requirements = 230V (-10%, + 6%), 16A three phase, 50/60Hz. Max permissible voltage between ground and neutral lines < 400mV.

 Compressed air supply = 50 to 75psig. (350 to 500Kpa).

 Reference gases = N2 and CO2.

 Environment = Heat output 2.5kW during normal operation; 4.5kW during system bake-out.

 Ambient temperature range = 64 to 82°F (18-28°C).

 Relative humidity range = 20 to 70%.

 Dimensions = 4.3ft H x 5.3ft W x 4.4ft D (130cm H x 160cm W x 134cm D).

 Weight = 1,430lbs (650kg).

Finnigan MAT 252 IRMS specs:

 Sensitivity = 800 - 1500 molecules CO2/mass 44 ion

 Internal precision 2s 13C @ 1 µl CF = 0.03ppt; @ 3 µl CF = 0.005ppt.

 Internal precision 2s 18O @ 4 µl CF = 0.02ppt; @ 10 µl CF 0.01ppt.

 Internal precision 2s 15N @ 20 µl CF = 0.010ppt.

 Internal precision 2s 32S @ 10 µl CF = 0.006ppt.

 External precision 1s 13C @ 1 µl CF = 0.1ppt; @ 3 µl CF = 0.01ppt.

 External precision 1s 18O @ 4 µl CF = 0.05ppt; @ 10 µl CF 0.03ppt.

 External precision 1s 15N @ 20 µl CF = 0.1ppt

 External precision 1s 32S @ 10 µl CF = 0.01ppt.

 Linearity 0.02 ppt/nA ion current mass 44


A Finnigan Kiel III carbonate device is interfaced with MAT 252 IRMS. For more information refer to the Kiel III specification sheet.