Finnigan MAT Delta plus Isotope Ratio Mass Spectrometer (IRMS):
Description: The Finnigan MAT Delta plus IRMS has a continuous flow (Conflo IITM) interface connecting an elemental analyzer, three faraday collectors, with a Pentium PC running ISODATTM software.
Power requirements = 230V (-10%, + 6%), 16A three phase, 50/60Hz. Max permissible voltage between ground and neutral lines < 400mV.
Compressed air supply = 50 to 75psig. (350 to 500Kpa).
Reference gases for N2 and CO2.
Environment = Heat output 2.5kW during normal operation; 4.5kW during system bake-out.
Ambient temperature range = 64 to 82°F (18-28°C).
Relative humidity range = 20 to 70%.
Dimensions = 33.5in H x 34.3in W x 36.6in D (85cm H x 87cm W x 93cm D).
Weight = 550lbs (250kg).
Finnigan MAT Delta plus IRMS specs:
Sensitivity = 1500 molecules CO2/mass 44 ion
Internal precision 2s 13C @ 3 µl CF = 0.020ppt; @ 10 µl CF = 0.006ppt.
Internal precision 2s 18O @ 5 µl CF = 0.04ppt; @ 10 µl CF = 0.012ppt.
Internal precision 2s 15N @ 10 µl CF = 0.05ppt; @ 20 µl CF = 0.010ppt.
Internal precision 2s 32S @ 10 µl CF = 0.010ppt.
External precision 1s 13C @ 3 µl CF = 0.05ppt; @ 10 µl CF = 0.02ppt.
External precision 1s 18O @ 5 µl CF = 0.10ppt; @ 10 µl CF 0.04ppt.
External precision 1s 15N @ 10 µl CF = 0.15ppt; @ 20 µl CF = 0.03ppt
External precision 1s 32S @ 10 µl CF = 0.02ppt.
Linearity 0.02 ppt/nA ion current mass 44
A Carlo-Erba Na1500 Series II elemental analyzer is connected to the Finnigan Delta plus IRMS. For more information refer to the
Carlo-Erba Na1500 Series II specification sheet.Department of Geological and Environmental Sciences / Stanford University
Last modified on 5/11/05 by David Mucciarone (e-mail address - dam@pangea.stanford.edu).