Carlo-Erba (CE Elantech, Inc.) Na1500 Series II elemental analyzer is connected to the Finnigan Delta plus IRMS via the Finnigan ConFlo II interface:
Power requirements: = 230V (-10%, + 6%), 16A three phase, 50/60Hz.
Compressed gas pressure requirements: Air, He, and O2 supply = 40 to 60psig. (300 to 400Kpa).
Reference gases: required for N2, CO2, and SO2.
Compressed gas requirement: UHP (4.6 He 99.996 purity) = 100ml/min @ 100Kpa.
Compressed gas requirement: UHP (4.3 O2 99.993 purity) = 20ml/min @ 100Kpa.
Ambient temperature range: = 64 to 82°F (18-28°C).
Relative humidity range: = 20 to 70%.
Dimensions: = 1.9ft H x 2.9ft W x 1.9ft D (58cm H x 88cm W x 57cm D).
Weight: = 154lbs (70kg).
Carlo Erba NA1500 Series 2 specs:
Sample amount: = 0.5 - 100mg.
Sample capacity: = 50 sample continuous automated run.
Reproducibility: = 0.1%
Detection limit: = 10ppm
Analysis time: N = 3 minutes, NC = 6 minutes, NCS/NCH = 10 minutes.
Measuring range: = 0.01 - 100%
Accuracy: At 100ppm in N, C, S mode = 0.1%